Question 1
A US hospital suspects a catheter contributed to a patient death. Under 21 CFR 803.30(a)(1), to whom must the user facility report?
Show official answer
Accepted answer: C
C is correct. Section 803.30(a)(1) requires user facilities to report deaths to FDA within 10 work days and also to the manufacturer when known. A is wrong because 30 calendar days is manufacturer MDR timing under §803.50(a), not user-facility death reporting.